Dayu Electronics: Applicability Analysis of Ultrasonic Sludge Level Meters in Flocculant-Dosed Conditions
When flocculant dosing creates abundant cotton-like flocs in water, whether an ultrasonic sludge level meter can be used must be judged separately by measuring distance and floc concentration. For short-distance conditions, optimized installation may be attempted; for long distances or high floc concentrations, ultrasonic solutions are generally not recommended.

Flocculent suspended matter significantly attenuates ultrasonic signals
When ultrasound propagates through water and encounters cotton-like flocs, suspended particles, or foam layers, absorption and scattering occur, causing substantial acoustic energy loss. At short distances of 1–2 meters, the round-trip path is short and the remaining echo signal may still be recognized by the receiving circuit; however, at 5–10 meters, the sound wave attenuates once going down and once returning, with overall signal attenuation reaching hundreds of thousands of times. The receiving end often cannot extract a valid echo, resulting in signal loss or data jumps.

Short-distance conditions may allow optimized installation
If the measuring distance is only 1–2 meters and the flocs do not completely block the probe's acoustic beam path, stability can be improved by installing in areas with fewer flocs, adding a waveguide tube, avoiding dosing ports and agitation zones, and regularly cleaning the probe. If floc concentration is high and a continuous floc or foam layer forms on the water surface, ultrasound may still fail.

For long distances or high concentrations, alternative solutions are recommended
When measuring distances exceed 5 meters and floc concentration is high, the reliability of ultrasonic sludge level meters drops significantly. For such conditions, alternatives such as radar level meters, pressure level meters, guided-wave or contact sludge level meters can be evaluated; the specific choice depends on tank structure, installation conditions, and control requirements.

On-site evaluation by the Dayu Electronics technical team is recommended
It is recommended to provide information such as measuring distance, floc concentration, whether a continuous floc layer forms on the water surface, tank depth, installation location, and control requirements. The Dayu Electronics technical team can determine whether an ultrasonic sludge level meter is suitable based on actual conditions, or recommend a more reliable alternative measurement solution.