Core Principle: Measuring Sound Velocity, Not Concentration Directly
The core function of the Dayu Electronics sound velocity sensor is high-precision measurement of liquid sound velocity, not direct output of alcohol concentration values. Alcohol concentration detection is based on the physical property that different concentrations of alcohol solutions correspond to different sound velocities, derived indirectly through sound velocity data conversion.

Concentration Accuracy Depends on Calibration, Not a Single Device Parameter
Regarding user concerns about accuracy, such as whether the error can reach 0.5% or 1%, it must be clarified: the sound velocity meter itself outputs raw sound velocity data, and the accuracy of concentration values depends entirely on the rigor of the calibration process. Users need to use standard alcohol solutions of known concentrations for multi-point calibration to establish a dedicated sound velocity-concentration correspondence curve or lookup table. The denser the calibration points and the more stable the environmental conditions, the higher the inversion accuracy.

Feasibility of 95% Alcohol Detection
Detecting 95% alcohol is completely feasible in principle, provided that the user has complete calibration capabilities. It is recommended that before actual application, a series of alcohol standard samples of known concentrations be prepared to cover the target detection range, and their sound velocity values be measured one by one to establish an accurate mapping relationship. The sound velocity sensor is responsible for providing high-repeatability, high-stability sound velocity data; concentration accuracy is the result of the combined effect of device performance and user calibration processes.

Key Interference Factors Affecting Accuracy
In actual measurements, temperature, bubbles, impurities, etc. will all interfere with sound velocity. If pursuing high precision at the 0.5% level, the temperature stability of the measurement environment must be strictly controlled, and the presence of bubbles or suspended particles in the liquid must be avoided. Dayu Electronics sound velocity meters provide a reliable sound velocity measurement foundation; the realization of final concentration accuracy requires users to simultaneously control conditions during calibration and use.